skip to main content
OSTI.GOV title logo U.S. Department of Energy
Office of Scientific and Technical Information

Title: Development of a correction factor for Xe-133 vials for use with a dose calibrator

Journal Article · · J. Nucl. Med.; (United States)
OSTI ID:5492542

Manufacturers of dose calibrators who give calibration settings for various radionuclies sometimes do not specify the type of radionuclide container the calibration is for. The container, moreover, may not be of the same type as those a user might purchase. When these factors are not considered, the activity administered to the patient may be significantly different from that intended. An experiment is described in which calibration factors are determined for measurement of Xe-133 activity in vials in a dose calibrator. This was accomplished by transferring the Xe-133 from the commercial vials to standard NBS calibration ampules. Based on ten such transfers, the resulting correction factor for the dose calibrator was 1.22.

Research Organization:
Univ. of Cincinnati Medical Center, OH
OSTI ID:
5492542
Journal Information:
J. Nucl. Med.; (United States), Vol. 23:4
Country of Publication:
United States
Language:
English