Rapid air titration method for determining SO/sub 2/ concentration in inhalation chambers

Journal Article · · Am. Ind. Hyg. Assoc. J.; (United States)
A rapid air titration method for determining SO/sub 2/ concentration in inhalation chambers has been validated using the pararosaniline-formaldehyde (PRA) method of West and Gaeke. This air-titration (iodate) method is an adaptation of iodometric methods using a starch indicator. Potassium iodate and an excess of potassium iodide are used in the reaction. Sampling is completed in ten minutes or less and concentration is calculated by use of a simple formula. Linear equations were derived over the range of concentrations from 0.5 to 100 ppm SO/sub 2/ for uncorrected iodate bubbler results, data corrected for tandem bubbler concentrations and data corrected for mean iodate bubbler efficiency. Linear correlation with the PRA method over this range was 0.999 for all three sets of data.