Quality control of scintillation cameras using a minicomputer
Journal Article
·
· J. Nucl. Med.; (United States)
OSTI ID:5466163
A minicomputer-based technique compiles objective indicators of scintigraphic system performance. The evaluation begins with the acquisition of a single image of an orthogonal hole pattern from which quantitative and regional measurements of point-source sensitivity, spatial resolution, and spatial linearity are derived. Two computer programs offer the user different but complementary features. The first program is the basis of an evaluation performed by a technologist for purposes of quality control. Operator intervention is minimal, and the entire protocol, including data acquisition and processing, can be completed in 20 min. The results are automatically compiled and displayed as graphs showing 100 consecutive sets of daily performance measurements. A second computer program is designed as an interactive diagnostic and research tool to display measurements as histograms and functional images. The operator can use the program to determine the quantitative and spatial characteristics of the system's intrinsic performance measurements made during the quality-control evaluations.
- Research Organization:
- Veterans Administration Medical Center, Denver, Colorado
- OSTI ID:
- 5466163
- Journal Information:
- J. Nucl. Med.; (United States), Journal Name: J. Nucl. Med.; (United States) Vol. 22:12; ISSN JNMEA
- Country of Publication:
- United States
- Language:
- English
Similar Records
Quality control of scintillation cameras using a minicomputer
Quality control for scintillation cameras. Final report
Quality control for scintillation cameras. [Compilation of multipurpose phantoms used in nuclear medicine]
Journal Article
·
Mon Nov 30 23:00:00 EST 1981
· J. Nucl. Med.; (United States)
·
OSTI ID:5797916
Quality control for scintillation cameras. Final report
Technical Report
·
Tue Jun 01 00:00:00 EDT 1976
·
OSTI ID:7292405
Quality control for scintillation cameras. [Compilation of multipurpose phantoms used in nuclear medicine]
Technical Report
·
Tue Jun 01 00:00:00 EDT 1976
·
OSTI ID:7267165
Related Subjects
550601* -- Medicine-- Unsealed Radionuclides in Diagnostics
62 RADIOLOGY AND NUCLEAR MEDICINE
COMPUTER CALCULATIONS
COMPUTERS
CONTROL
COUNTING TECHNIQUES
DATA ACQUISITION
DATA PROCESSING
DIAGNOSTIC TECHNIQUES
ELECTRONIC CIRCUITS
MICROELECTRONIC CIRCUITS
MICROPROCESSORS
PROCESSING
QUALITY CONTROL
RADIOISOTOPE SCANNING
RESOLUTION
SCINTISCANNING
SPATIAL RESOLUTION
62 RADIOLOGY AND NUCLEAR MEDICINE
COMPUTER CALCULATIONS
COMPUTERS
CONTROL
COUNTING TECHNIQUES
DATA ACQUISITION
DATA PROCESSING
DIAGNOSTIC TECHNIQUES
ELECTRONIC CIRCUITS
MICROELECTRONIC CIRCUITS
MICROPROCESSORS
PROCESSING
QUALITY CONTROL
RADIOISOTOPE SCANNING
RESOLUTION
SCINTISCANNING
SPATIAL RESOLUTION