To determine the usefulness of the rotating slant-hole (RSH) collimator, a new imaging device for tomographic myocardial imaging, we evaluated its performance in phantom studies. Perfusion defects of variable size, location, and tracer concentration were simulated and imaging was performed with a conventional parallel-hole collimator and the RSH collimator. Planar and depth resolutions assessed with a line source compared favorably with those reported previously for the seven-pinhole collimator. The results also showed RSH tomography to be more sensitive than planar imaging in detecting simulated myocardial defects but indicated some limitations, especially the occurrence of artifactual defects that may reduce the specificity of the imaging device for the detection of perfusion abnormalities.
- Research Organization:
- UCLA School of Medicine, Los Angeles
- OSTI ID:
- 5450851
- Journal Information:
- J. Nucl. Med.; (United States), Journal Name: J. Nucl. Med.; (United States) Vol. 23:1; ISSN JNMEA
- Country of Publication:
- United States
- Language:
- English
Similar Records
Performance of the rotating slant-hole collimator for the detection of myocardial perfusion abnormalities
A new collimator for cardiac tomography: the quadrant slant-hole collimator
I-123 HIPDM brain imaging with a rotating gamma camera and slant-hole collimator
Journal Article
·
Thu Dec 31 23:00:00 EST 1981
· J. Nucl. Med.; (United States)
·
OSTI ID:5638375
A new collimator for cardiac tomography: the quadrant slant-hole collimator
Journal Article
·
Wed Sep 01 00:00:00 EDT 1982
· J. Nucl. Med.; (United States)
·
OSTI ID:6656833
I-123 HIPDM brain imaging with a rotating gamma camera and slant-hole collimator
Journal Article
·
Sat Mar 31 23:00:00 EST 1984
· J. Nucl. Med.; (United States)
·
OSTI ID:6879911
Related Subjects
440300 -- Miscellaneous Instruments-- (-1989)
47 OTHER INSTRUMENTATION
550601* -- Medicine-- Unsealed Radionuclides in Diagnostics
550901 -- Pathology-- Tracer Techniques
59 BASIC BIOLOGICAL SCIENCES
62 RADIOLOGY AND NUCLEAR MEDICINE
ANIMAL TISSUES
BODY
CARDIOVASCULAR SYSTEM
COLLIMATORS
COMPARATIVE EVALUATIONS
COUNTING TECHNIQUES
DIAGNOSIS
DIAGNOSTIC TECHNIQUES
DIAGNOSTIC USES
HEART
IMAGES
ISOTOPES
MALFORMATIONS
MOCKUP
MUSCLES
MYOCARDIUM
ORGANS
PATHOLOGICAL CHANGES
PATHOLOGY
PERFORMANCE TESTING
PERFUSED TISSUES
PHANTOMS
RADIOISOTOPE SCANNING
RADIOISOTOPES
RESOLUTION
SCINTISCANNING
STRUCTURAL MODELS
TESTING
TISSUES
TOMOGRAPHY
USES
47 OTHER INSTRUMENTATION
550601* -- Medicine-- Unsealed Radionuclides in Diagnostics
550901 -- Pathology-- Tracer Techniques
59 BASIC BIOLOGICAL SCIENCES
62 RADIOLOGY AND NUCLEAR MEDICINE
ANIMAL TISSUES
BODY
CARDIOVASCULAR SYSTEM
COLLIMATORS
COMPARATIVE EVALUATIONS
COUNTING TECHNIQUES
DIAGNOSIS
DIAGNOSTIC TECHNIQUES
DIAGNOSTIC USES
HEART
IMAGES
ISOTOPES
MALFORMATIONS
MOCKUP
MUSCLES
MYOCARDIUM
ORGANS
PATHOLOGICAL CHANGES
PATHOLOGY
PERFORMANCE TESTING
PERFUSED TISSUES
PHANTOMS
RADIOISOTOPE SCANNING
RADIOISOTOPES
RESOLUTION
SCINTISCANNING
STRUCTURAL MODELS
TESTING
TISSUES
TOMOGRAPHY
USES