Skip to main content
U.S. Department of Energy
Office of Scientific and Technical Information

Graphite calorimeter in water phantom and calibration of ionization chambers in dose to water for /sup 60/Co gamma radiation

Journal Article · · Med. Phys.; (United States)
OSTI ID:5446290
A graphite calorimeter similar in size and shape to a Farmer type 0.6-cm/sup 3/ ionization chamber was developed for direct measurement of absorbed dose at a specified depth in a water phantom irradiated by a beam of /sup 60/Co gamma radiation. The accuracy of the absorbed dose determined was estimated to be +- 1.1% at a dose level of 4 rad/s (4 x 10/sup -2/Gy s/sup -1/). The absorbed dose to water at the calibration depth of 5 cm was standardized for our /sup 60/Co gamma beam, for a 60-cm SSD and 10 x 10 cm/sup 2/ field. Farmer type ionization chambers were calibrated against the standardized absorbed dose. The overall accuracy in calibrating the ionization chambers was +- 1.2%.
Research Organization:
Radiological Standards Laboratory, Division of Radiological Protection, Bhabha Atomic Research Centre, Trombay, Bombay-400 085, India
OSTI ID:
5446290
Journal Information:
Med. Phys.; (United States), Journal Name: Med. Phys.; (United States) Vol. 7:3; ISSN MPHYA
Country of Publication:
United States
Language:
English