High-spatial-resolution MWPC for medical imaging with positron emitters
Conference
·
OSTI ID:5430820
The performance of a 15 x 15 cm/sup 2/ MultiWire Proportional Chamber (MWPC) with lead glass tube converter is presented. A standard Argon-Methane (70-30) gas mixture has been used at various pressures between 1.0 and 2.0 atm. For 511 keV ..gamma..-rays incident almost perpendicular onto a 1.0 cm thick lead glass tube converter, a detection efficiency of 4.5%, a time resolution of 130 ns (FWHM) and a spatial resolution of 1.3 mm (FWHM) has been obtained at 2 atm. Fast delay lines, capacitively coupled to the cathodes, standard integrated amplifiers and comparator electronics have been adopted for the position read-out. The use of fast delay lines (specific delay 8 ns/cm) ensures a high rate capability. A multiplanar positron camera is proposed, made of six MWPC modules arranged to form the lateral surface of a hexagonal prism. Each module (50 x 50 cm/sup 2/) will have a 2 cm thick converter on both sides of the chamber and will be operating at 2 atm. The expected performance is presented and compared with that of a recent multi-ring BGO camera, and is shown to be superior in spatial resolution and comparable in event rate at a considerably lower cost.
- Research Organization:
- Lawrence Berkeley Lab., CA (USA)
- DOE Contract Number:
- W-7405-ENG-48
- OSTI ID:
- 5430820
- Report Number(s):
- LBL-14044; CONF-820354-2; ON: DE82014477
- Country of Publication:
- United States
- Language:
- English
Similar Records
Medical positron imaging with a dense drift space multiwire proportional chamber
Design considerations for a high spatial resolution positron camera with dense drift space MWPC's
Design considerations for a high-spatial-resolution positron camera with dense-drift-space MWPC's
Journal Article
·
Thu Jul 01 00:00:00 EDT 1982
· IEEE Transactions on Medical Imaging; (United States)
·
OSTI ID:5326078
Design considerations for a high spatial resolution positron camera with dense drift space MWPC's
Journal Article
·
Mon Jan 31 23:00:00 EST 1983
· IEEE Trans. Nucl. Sci.; (United States)
·
OSTI ID:6155718
Design considerations for a high-spatial-resolution positron camera with dense-drift-space MWPC's
Conference
·
Fri Oct 01 00:00:00 EDT 1982
·
OSTI ID:6944886
Related Subjects
440101* -- Radiation Instrumentation-- General Detectors or Monitors & Radiometric Instruments
46 INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY
550601 -- Medicine-- Unsealed Radionuclides in Diagnostics
62 RADIOLOGY AND NUCLEAR MEDICINE
CAMERAS
DESIGN
EFFICIENCY
GAMMA CAMERAS
MEASURING INSTRUMENTS
MULTIWIRE PROPORTIONAL CHAMBERS
PERFORMANCE
POSITRON CAMERAS
PROPORTIONAL COUNTERS
RADIATION DETECTORS
RESOLUTION
SPATIAL RESOLUTION
46 INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY
550601 -- Medicine-- Unsealed Radionuclides in Diagnostics
62 RADIOLOGY AND NUCLEAR MEDICINE
CAMERAS
DESIGN
EFFICIENCY
GAMMA CAMERAS
MEASURING INSTRUMENTS
MULTIWIRE PROPORTIONAL CHAMBERS
PERFORMANCE
POSITRON CAMERAS
PROPORTIONAL COUNTERS
RADIATION DETECTORS
RESOLUTION
SPATIAL RESOLUTION