skip to main content
OSTI.GOV title logo U.S. Department of Energy
Office of Scientific and Technical Information

Title: Corrosion and degradation of a polyurethane/Co-Ni-Cr-Mo pacemaker lead

Journal Article · · J. Biomed. Mater. Res.; (United States)
OSTI ID:5428047

An investigation to study changes in the metal surfaces and the polyurethane insulation of heart pacemaker leads under controlled in vitro conditions was conducted. A polyurethane (Pellethane 2363-80A)/Co-Ni-Cr-Mo (MP35N) wire lead was exposed in Hanks' physiological saline solution for 14 months and then analyzed using scanning electron microscopy, x-ray energy dispersive analysis, and small angle x-ray scattering. Results showed that some leakage of solution into the lead had occurred and changes were present on both the metal and the polyurethane surfaces.

Research Organization:
Center for Devices and Radiological Health, Food and Drug Administration, Rockville, MD (USA)
OSTI ID:
5428047
Journal Information:
J. Biomed. Mater. Res.; (United States), Vol. 21
Country of Publication:
United States
Language:
English