Reconstruction of 4-MV bremsstrahlung spectra from measured transmission data
Journal Article
·
· Med. Phys.; (United States)
Transmission data for 4-MV bremsstrahlung beams have been measured with a combination of lead and aluminum attenuators. From these data, the original energy spectra have been reconstructed using an iterative least-squares technique, previously evaluated by simulation studies. The spectra on the central axis for three similar 4-MV linear accelerators indicated no significant differences. When studying the spectra at 5/sup 0/ and 9/sup 0/ off the central axis, that at 9/sup 0/ showed the expected increase of low-energy photons. All these spectra showed a maximum photon energy of 4.5 +- 0.2 MeV. When the magnetron power was reduced, the spectrum on the central axis shifted to lower energies and the maximum photon energy decreased to 3.5 +- 0.2 MeV. The result of this experimental study confirms the conclusions from the previous simulation, that the numerical technique for analysis of transmission data can accurately represent 4-MV bremsstrahlung spectra and detect differences in energy distribution with changes in machine tuning and position in the radiation field for a 4-MV bremsstrahlung beam.
- Research Organization:
- Joint Center for Radiation Therapy and Department of Radiation Therapy, Harvard Medical School, Boston, Massachusetts 02115
- OSTI ID:
- 5385964
- Journal Information:
- Med. Phys.; (United States), Journal Name: Med. Phys.; (United States) Vol. 10:6; ISSN MPHYA
- Country of Publication:
- United States
- Language:
- English
Similar Records
Characteristics of the high-energy photon beam of a 25-MeV accelerator
A study of quality of bremsstrahlung spectra reconstructed from transmission measurements
Large efficiency improvements in BEAMnrc using directional bremsstrahlung splitting
Journal Article
·
Thu Dec 31 23:00:00 EST 1987
· Med. Phys.; (United States)
·
OSTI ID:5732011
A study of quality of bremsstrahlung spectra reconstructed from transmission measurements
Journal Article
·
Wed Feb 28 23:00:00 EST 1990
· Medical Physics; (USA)
·
OSTI ID:6765791
Large efficiency improvements in BEAMnrc using directional bremsstrahlung splitting
Journal Article
·
Fri Oct 01 00:00:00 EDT 2004
· Medical Physics
·
OSTI ID:20634606
Related Subjects
654001* -- Radiation & Shielding Physics-- Radiation Physics
Shielding Calculations & Experiments
73 NUCLEAR PHYSICS AND RADIATION PHYSICS
ALUMINIUM
ATTENUATION
BIOMEDICAL RADIOGRAPHY
BREMSSTRAHLUNG
DIAGNOSTIC TECHNIQUES
DOSIMETRY
ELECTROMAGNETIC RADIATION
ELEMENTS
ENERGY DEPENDENCE
IONIZATION CHAMBERS
IONIZING RADIATIONS
LEAD
LEAST SQUARE FIT
MAXIMUM-LIKELIHOOD FIT
MEASURING INSTRUMENTS
MEDICINE
METALS
NEUTRAL-PARTICLE TRANSPORT
NUCLEAR MEDICINE
NUMERICAL SOLUTION
PHOTON TRANSPORT
RADIATION DETECTORS
RADIATION TRANSPORT
RADIATIONS
RADIOLOGY
SPECTRA
TRANSMISSION
X RADIATION
X-RAY DOSIMETRY
Shielding Calculations & Experiments
73 NUCLEAR PHYSICS AND RADIATION PHYSICS
ALUMINIUM
ATTENUATION
BIOMEDICAL RADIOGRAPHY
BREMSSTRAHLUNG
DIAGNOSTIC TECHNIQUES
DOSIMETRY
ELECTROMAGNETIC RADIATION
ELEMENTS
ENERGY DEPENDENCE
IONIZATION CHAMBERS
IONIZING RADIATIONS
LEAD
LEAST SQUARE FIT
MAXIMUM-LIKELIHOOD FIT
MEASURING INSTRUMENTS
MEDICINE
METALS
NEUTRAL-PARTICLE TRANSPORT
NUCLEAR MEDICINE
NUMERICAL SOLUTION
PHOTON TRANSPORT
RADIATION DETECTORS
RADIATION TRANSPORT
RADIATIONS
RADIOLOGY
SPECTRA
TRANSMISSION
X RADIATION
X-RAY DOSIMETRY