The development, characterization, and performance evaluation of a new combination type personnel neutron dosimeter
- Oak Ridge National Lab. (ORNL), Oak Ridge, TN (United States)
- Texas A and M Univ., College Station, TX (United States). Dept. of Nuclear Engineering
A new combination type personnel neutron dosimeter has been designed and developed at the Oak Ridge National Laboratory (ORNL). The combination personnel neutron dosimeter (CPND) consists of a Harshaw albedo neutron thermoluminescent dosimeter (two pairs of TLD-600/TLD- 700) and two bubble detectors (one BD-100R and one BDS-1500 from Bubble Technology Industries, Canada). The CPND was developed with the aim of having crude neutron spectrometric capability, universal applicability, better angular response, and an improved lower limit of detection (LLD). The CPND has been well characterized in the following areas: reusability, linearity, lower limit of detection (LLD), detection capability in mixed neutron-gamma fields, angular dependence, and neutron energy dependence. the characterization was accomplished with irradiations using a {sup 238}Pu-Be source, a {sup 252}Cf(D{sub 2}O) source, a {sup 252}Cf source, a {sup 252}Cf(PE) source, monoenergetic neutrons from accelerator and reactor filtered beams, {sup 137}Cs, and X-rays. Optimum signal readout procedures, signal processing techniques, routine operational usage, and neutron dose equivalent evaluation algorithms for the CPND were developed with the goals of having the best precision and accuracy as well as being convenient to use. 97 refs., 43 figs., 22 tabs.
- Research Organization:
- Oak Ridge National Lab., TN (USA)
- Sponsoring Organization:
- DOE/ER
- DOE Contract Number:
- AC05-84OR21400
- OSTI ID:
- 5364283
- Report Number(s):
- ORNL-6593; ON: DE90004408
- Country of Publication:
- United States
- Language:
- English
Similar Records
The development, characterization, and performance evaluation of a new combination type personnel neutron dosimeter
Site-specific calibration of the Hanford personnel neutron dosimeter
Improving neutron dosimetry using bubble detector technology
Thesis/Dissertation
·
Sat Dec 31 23:00:00 EST 1988
·
OSTI ID:6148054
Site-specific calibration of the Hanford personnel neutron dosimeter
Conference
·
Sat Oct 01 00:00:00 EDT 1994
·
OSTI ID:10107319
Improving neutron dosimetry using bubble detector technology
Technical Report
·
Sun Jan 31 23:00:00 EST 1993
·
OSTI ID:10152596
Related Subjects
440102* -- Radiation Instrumentation-- Radiation Dosemeters
46 INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY
DESIGN
DOSEMETERS
DOSIMETRY
EVALUATION
MEASURING INSTRUMENTS
MONITORING
PERFORMANCE TESTING
PERSONNEL DOSIMETRY
PERSONNEL MONITORING
RADIATION MONITORING
RADIATION PROTECTION
TESTING
THERMOLUMINESCENT DOSIMETRY
46 INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY
DESIGN
DOSEMETERS
DOSIMETRY
EVALUATION
MEASURING INSTRUMENTS
MONITORING
PERFORMANCE TESTING
PERSONNEL DOSIMETRY
PERSONNEL MONITORING
RADIATION MONITORING
RADIATION PROTECTION
TESTING
THERMOLUMINESCENT DOSIMETRY