Subpicosecond x-ray streak camera

Conference · · Proceedings of SPIE - The International Society for Optical Engineering
DOI:https://doi.org/10.1117/12.50528· OSTI ID:5320759
 [1];  [1];  [1]
  1. Lawrence Livermore National Laboratory (LLNL), Livermore, CA (United States)

We have designed and built an x-ray streak camera with subpicosecond time resolution. This camera attains its fast temporal resolution through a very strong extraction field, 100,000 V/cm, at the photocathode. It incorporates a narrow electron emission band photocathode that will also help the time resolution. The total time resolution has been calculated to be near 600 fs.

Research Organization:
Lawrence Livermore National Laboratory (LLNL), Livermore, CA (United States)
Sponsoring Organization:
USDOE
DOE Contract Number:
W-7405-ENG-48
OSTI ID:
5320759
Report Number(s):
UCRL-JC--107984; CONF-9107115--38; ON: DE91017664
Journal Information:
Proceedings of SPIE - The International Society for Optical Engineering, Journal Name: Proceedings of SPIE - The International Society for Optical Engineering Vol. 1539; ISSN 0277-786X
Publisher:
SPIE
Country of Publication:
United States
Language:
English

Similar Records

Characteristics of an ultrafast x-ray streak camera
Conference · Wed Jun 01 00:00:00 EDT 1994 · OSTI ID:10196773

Ultrafast x-ray streak camera for use in ultrashort laser-produced plasma research
Journal Article · Sat Dec 31 23:00:00 EST 1994 · Review of Scientific Instruments; (United States) · OSTI ID:6700470

An x-ray streak camera with high spatio-temporal resolution
Journal Article · Mon Sep 24 00:00:00 EDT 2007 · Applied Physics Letters · OSTI ID:21013661