Subpicosecond x-ray streak camera
Conference
·
· Proceedings of SPIE - The International Society for Optical Engineering
- Lawrence Livermore National Laboratory (LLNL), Livermore, CA (United States)
We have designed and built an x-ray streak camera with subpicosecond time resolution. This camera attains its fast temporal resolution through a very strong extraction field, 100,000 V/cm, at the photocathode. It incorporates a narrow electron emission band photocathode that will also help the time resolution. The total time resolution has been calculated to be near 600 fs.
- Research Organization:
- Lawrence Livermore National Laboratory (LLNL), Livermore, CA (United States)
- Sponsoring Organization:
- USDOE
- DOE Contract Number:
- W-7405-ENG-48
- OSTI ID:
- 5320759
- Report Number(s):
- UCRL-JC--107984; CONF-9107115--38; ON: DE91017664
- Journal Information:
- Proceedings of SPIE - The International Society for Optical Engineering, Journal Name: Proceedings of SPIE - The International Society for Optical Engineering Vol. 1539; ISSN 0277-786X
- Publisher:
- SPIE
- Country of Publication:
- United States
- Language:
- English
Similar Records
Characteristics of an ultrafast x-ray streak camera
Ultrafast x-ray streak camera for use in ultrashort laser-produced plasma research
An x-ray streak camera with high spatio-temporal resolution
Conference
·
Wed Jun 01 00:00:00 EDT 1994
·
OSTI ID:10196773
Ultrafast x-ray streak camera for use in ultrashort laser-produced plasma research
Journal Article
·
Sat Dec 31 23:00:00 EST 1994
· Review of Scientific Instruments; (United States)
·
OSTI ID:6700470
An x-ray streak camera with high spatio-temporal resolution
Journal Article
·
Mon Sep 24 00:00:00 EDT 2007
· Applied Physics Letters
·
OSTI ID:21013661