Pulsed dye-laser excitation with multichannel scaler photon counting is used to obtain time resolved emission spectra of uranyl ions in aqueous solution. Kinetic analysis of this data corrects for matrix quenching and temperature effects which reduce the quantum yield of the uranyl ion luminescence. The method gives accurate measurements without separative prechemistry or the use of internal standards. Detection limits of one part-per-trillion (pptr) have been demonstrated, and in samples with concentrations greater than 100 pptr, relative standard deviations of less than 3% are achieved routinely.