Technique of range modulation for a clinical pi-meson beam

Journal Article · · Rev. Sci. Instrum.; (United States)
DOI:https://doi.org/10.1063/1.1137089· OSTI ID:5295820
A simple, versatile technique of range modulation has been developed based on the mathematical technique of linear programming for linear addition of depth dose curves. This technique enables the use of discrete step range shifting, and hence results in substantial simplification in the design and operation of range modulators. A prototype using only eight steps has been constructed for the TRIUMF biomedical ..pi../sup -/ channel for cancer radiotherapy. This system has been shown to be capable of producing depth dose profiles of various characteristics required for biomedical experimentation. Its operation under computer control is also simple, accurate, and immune to large fluctuations in the output of the TRIUMF accelerator.
Research Organization:
Batho Biomedical Facility, TRIUMF, University of British Columbia, Vancouver, British Columbia, Canada V6T 2A3
OSTI ID:
5295820
Journal Information:
Rev. Sci. Instrum.; (United States), Journal Name: Rev. Sci. Instrum.; (United States) Vol. 53:7; ISSN RSINA
Country of Publication:
United States
Language:
English

Similar Records

Beam scanning system for a clinical pi-meson beam
Journal Article · Fri Feb 28 23:00:00 EST 1986 · Rev. Sci. Instrum.; (United States) · OSTI ID:6288871

Pi-meson range shifter for clinical therapy
Journal Article · Fri Dec 31 23:00:00 EST 1976 · Rev. Sci. Instrum.; (United States) · OSTI ID:7135205

Preparatory clinical studies of Pi-mesons at TRIUMF
Journal Article · Mon Dec 31 23:00:00 EST 1984 · Radiat. Res., Suppl.; (United States) · OSTI ID:5203595