Skip to main content
U.S. Department of Energy
Office of Scientific and Technical Information

Formation of homogeneous field in a shielded superconducting solenoid for an MRI

Conference · · IEEE Transactions on Magnetics (Institute of Electrical and Electronics Engineers); (United States)
OSTI ID:5294404
; ; ; ; ;  [1]
  1. D.V. Efremov Scientific Research Inst. of Electrophysical Apparatus, Leningrad (SU)
This paper proposes methods for homogenizing and shimming a high homogeneity solenoid with a magnetic shield. The methods make the shield design more accurate and the shim coils less complicated. The whole-body 0.5 T superconducting MRI magnet is designed to achieve a field homogeneity of 3 ppm in a 50 cm DSV. Its fringe field is less than 0.5 mT outside a radius of 5 m using 12 tons of iron.
OSTI ID:
5294404
Report Number(s):
CONF-910662--
Conference Information:
Journal Name: IEEE Transactions on Magnetics (Institute of Electrical and Electronics Engineers); (United States) Journal Volume: 28:1
Country of Publication:
United States
Language:
English