Skip to main content
U.S. Department of Energy
Office of Scientific and Technical Information

Comments on geometric deconvolution: a meta-algorithm for limited view computed tomography

Journal Article · · IEEE Trans. Bio-Med. Eng.; (United States)
A recent paper on improving computed tomography images suffering from artifacts due to a missing sector of projection angles is criticized. It is shown that the proposed deconvolution approach is not applicable, since the transfer function to be inverted is zero in the whole sector of interest.
Research Organization:
Technische Universitaet Muenchen, West Germany
OSTI ID:
5277363
Journal Information:
IEEE Trans. Bio-Med. Eng.; (United States), Journal Name: IEEE Trans. Bio-Med. Eng.; (United States) Journal Issue: 3 Vol. BME-32:3; ISSN IEBEA
Country of Publication:
United States
Language:
English

Similar Records

Reply to comments on geometric deconvolution: a meta-algorithm for limited view computed tomography
Journal Article · Thu Feb 28 23:00:00 EST 1985 · IEEE Trans. Bio-Med. Eng.; (United States) · OSTI ID:5277364

Image restoration by Wiener deconvolution in limited-view computed tomography
Conference · Sat Nov 30 23:00:00 EST 1985 · Appl. Opt.; (United States) · OSTI ID:6698597

Quantitative and qualitative evaluation of geometric deconvolution of distortion in limited-view computed tomography
Journal Article · Wed May 01 00:00:00 EDT 1985 · IEEE Trans. Bio-Med. Eng.; (United States) · OSTI ID:5277377