Skip to main content
U.S. Department of Energy
Office of Scientific and Technical Information

Gamma-ray measurements for simultaneous calorimetric assay

Technical Report ·
DOI:https://doi.org/10.2172/5263121· OSTI ID:5263121
Gamma-ray measurements obtained in the course of developing a simultaneous calorimetric assay system are described. Gamma-ray measurements of the isotopic composition of six, well-characterized plutonium oxide samples were obtained while the samples were in the calorimeter. These samples represent a range of plutonium masses from 19 to 231 g and two isotopic compositions. The values of effective specific power determined from the gamma-ray measurements agree with the values determined from destructive assay.
Research Organization:
Mound Facility, Miamisburg, OH (USA)
DOE Contract Number:
AC04-76DP00053
OSTI ID:
5263121
Report Number(s):
MLM-2898; ON: DE82008615
Country of Publication:
United States
Language:
English