Title: Multi step parallel plate avalanche chamber as a 2D imager for MeV pulsed radiography

Book ·
OSTI ID:524743
 [1]; ;  [2]
  1. C.E.A., Courtry (France)
  2. ESPCI, Paris (France)

In this paper, the authors present the experimental validation of the wide signal measurement range (>100) of a high sensitivity gateable camera based on a Multi Step Parallel Plate Avalanche Chamber for MeV Pulsed Radiography applications. The MSPPAC was tested with a metallic converter. Thus, they have demonstrated that the Multi Step Parallel Plate Avalanche Chamber (atmospheric pressure) is a potentially high-performance imager, even for the high hard X-ray dose rates used in radiography and radiotherapy.

OSTI ID:
524743
Report Number(s):
CONF-960848-; ISBN 0-8194-2247-9; TRN: 97:016734
Resource Relation:
Conference: Denver `96: 1. conference on space processing of materials, at SPIE International Society for Optical Engineering (SPIE) annual international symposium on optical science, engineering, and instrumentation, Denver, CO (United States), 4-9 Aug 1996; Other Information: PBD: 1996; Related Information: Is Part Of Hard x-ray/gamma-ray and neutron optics, sensors, and applications; Hoover, R.B.; Doty, F.P. [eds.]; PB: 331 p.; Proceedings/SPIE, Volume 2859