Soil gas sample analysis method evaluation and comparison

Conference ·
OSTI ID:524679
; ;  [1]
  1. O`Brien and Gere Engineers, Inc., Syracuse, NY (United States)

The quality of soil gas survey data is a function of many variables, particularly the analytical method used to analyze the collected samples of VOCs. Several analytical methods are commonly employed, including direct reading photoionization detectors, portable gas chromatographs (GCs), and mobile laboratories equipped with benchtop GCs. Two analytical methods employing Hewlett-Packard (HP) 5890 Series 2 GCs were compared relative to their sensitivity, precision, accuracy, practicality, and productivity. The first method utilized direct injection as the method of sample introduction, and the second, a Tekmar LSC 2000/ALS 2016 purge and trap autosampler. The direct injection method provided greater accuracy, precision, and ease of use, while the purge and trap autosampler method provided increased sensitivity and higher productivity.

OSTI ID:
524679
Report Number(s):
CONF-9504292--; ISBN 0-8031-2043-5
Country of Publication:
United States
Language:
English