Radiac calibrators. Final report on test operations procedure
This test operations procedure (TOP) sets forth general and specific test procedures for measuring the technical performance of Radiac Calibrators. A comparison of the test results with the criteria of the technical specifications permits an evaluation of their suitability as Radiac Calibrators. This TOP applies to Radiac Calibrators designed for field-use only. This TOP describes three tests: Characteristic Calibration Energy, which characterizes the test item's radiation energy in a test cavity; Calibration Accuracy, which determines whether the test item can, with the required accuracy, calibrate each of its anticipated types of radiac meters; and Repeatability, which determines whether the test item can, with the same accuracy, repeat the required calibrations.
- Research Organization:
- Army Test and Evaluation Command, Aberdeen Proving Ground, MD (USA)
- OSTI ID:
- 5246285
- Report Number(s):
- AD-A-134611/3; TOP-8-2-064
- Country of Publication: