Skip to main content
U.S. Department of Energy
Office of Scientific and Technical Information

Pocket neutron REM meter

Conference ·
OSTI ID:5219535
This paper describes a pocket-calculator-sized, neutron-sensitive, REM-responding personnel dosimeter that uses three tissue-equivalent cylindrical proportional counters as neutron-sensitive detectors. These are conventionally called Linear Energy Transfer (LET) counters. Miniaturized hybrid circuits are used for the linear pulse handling electronics, followed by a 256-channel ADC. A CMOS microprocessor is used to calculate REM exposure from the basic rads-tissue data supplied by the LET counters and also to provide timing and display functions. The instrument is used to continuously accumulate time in hours since reset, total counts accumulated, rads-tissue, and REM. At any time the user can display any one of these items or a channel number (an aid in calibration). The instrument provides such data with a precision of +- 3% for a total exposure of 1 mREM over 8 hours.
Research Organization:
EG and G, Inc., Goleta, CA (USA)
DOE Contract Number:
AC08-76NV01183
OSTI ID:
5219535
Report Number(s):
CONF-820668-1; ON: DE82016694
Country of Publication:
United States
Language:
English

Similar Records

Pocket neutron REM meter
Conference · Wed Dec 31 23:00:00 EST 1980 · OSTI ID:5253336

Pocket neutron REM meter
Journal Article · Mon Jan 31 23:00:00 EST 1983 · IEEE Trans. Nucl. Sci.; (United States) · OSTI ID:5913397

Pocket neutron REM meter
Conference · Thu Dec 31 23:00:00 EST 1981 · OSTI ID:6794237