Skip to main content
U.S. Department of Energy
Office of Scientific and Technical Information

Radiation exposure reduction using carbon fiber

Journal Article · · Radiology; (United States)
The ability of carbon fiber cassette fronts to reduce x-ray exposure was evaluated by comparing a carbon fiber cassette with a Kodak X-Omatic cassette. Rare-earth screens were used. Although the transmission of the carbon fiber was considerably higher than that of the X-Omatic cassette, the radiation exposure reduction was significant only in the lower kVp ranges. Although some decrease in the primary-to-scatter ratio was measured, the overall image contrast with the carbon fiber cassette was only minimally reduced in the kVp range investigated.
Research Organization:
Department of Radiology, University of Miami School of Medicine, Miami, FL
OSTI ID:
5210483
Journal Information:
Radiology; (United States), Journal Name: Radiology; (United States) Vol. 142:3; ISSN RADLA
Country of Publication:
United States
Language:
English

Similar Records

Radiation exposure reduction using carbon fiber
Journal Article · Sun Feb 28 23:00:00 EST 1982 · Radiology; (United States) · OSTI ID:6931607

Effects of rare-earth filters on patient exposure and image contrast
Journal Article · Sat Aug 01 00:00:00 EDT 1987 · J. Dent. Res.; (United States) · OSTI ID:5976085

Radiation exposure reduction by use of Kevlar cassettes in the neonatal nursery
Journal Article · Fri May 01 00:00:00 EDT 1987 · AJR, Am. J. Roentgenol.; (United States) · OSTI ID:6352228