Skip to main content
U.S. Department of Energy
Office of Scientific and Technical Information

Linearity of quantized voltages generated by a Josephson-junction array

Journal Article · · IEEE (Institute of Electrical and Electronics Engineers) Transactions on Instrumentation and Measurement; (USA)
DOI:https://doi.org/10.1109/19.192294· OSTI ID:5201834
; ; ;  [1]
  1. Electrotechnical Lab., 1-1-4 Umezono, Tsukuba-shi, Ibaraki 305 (JP)

A measurement system is proposed to examine experimentally the linearity of quantized voltages generated by a Josephson-junction array at 100 MV and 1 V with an accuracy better than 1 X10/sup -8/. Ten resistors, connected in series, are driven by a constant current source and the total voltage drop is compared with the sum of the voltages across each resistor. A preliminary experiment at the 0.2-ppm level was performed. The largest problem for the lower accuracy than expected was the instability of the junction array's voltages, which seems to be attributable to imperfect cryostat design.

OSTI ID:
5201834
Journal Information:
IEEE (Institute of Electrical and Electronics Engineers) Transactions on Instrumentation and Measurement; (USA), Journal Name: IEEE (Institute of Electrical and Electronics Engineers) Transactions on Instrumentation and Measurement; (USA) Vol. 38:2; ISSN IEIMA; ISSN 0018-9456
Country of Publication:
United States
Language:
English