  2. Rochester Gas and Electric Corp., NY (United States)

This paper reports that Groups of Panasonic UD-802 thermoluminescent dosimeters (TLDs) were irradiated to successively increasing doses of Cesium-137 gamma radiation (0.662 MeV gamma rays) and then processed using a Panasonic UD-710 automatic TLD reader. The results were subjected to statistical tests to determine the critical level, the level of detection, and the less-than level. The critical level is equivalent to 1.7 mrad, the lower limit of detection is equivalent to 5 mrad and the less-than level has a high range value of 7.5 mrad.