New technique to improve the accuracy of albedo neutron dosimeter evaluations

Conference ·
OSTI ID:5169080
The calibration factor for albedo neutron dosimeters varies greatly depending upon the energy of the neutrons in the exposure. Calibration results obtained over an eight-year period at each Lawrence Livermore National Laboratory facility where neutron exposure may occur were reviewed. A stronger relationship than expected was found between the ratio of the readings of the 9-in. to 3-in. spheres and the percent thermal. Readings from personnel and albedo badges were reviewed. The readings were consistent with the use of a calibration factor for the albedo dosimeter which varies with changes in the ratio of the personnel and albedo dosimeter TLD readings. 2 references, 6 figures. (ACR)
Research Organization:
Lawrence Livermore National Lab., CA (USA)
DOE Contract Number:
W-7405-ENG-48
OSTI ID:
5169080
Report Number(s):
UCRL-88391; CONF-840512-6; ON: DE84002458