skip to main content
OSTI.GOV title logo U.S. Department of Energy
Office of Scientific and Technical Information

Title: Plutonium isotopic analysis of highly enriched mixed oxides

Technical Report ·
DOI:https://doi.org/10.2172/5162039· OSTI ID:5162039

We investigated the analysis method used by the International Atomic Energy Agency (IAEA) to determine the plutonium isotopic composition of highly enriched mixed oxides (MOX). The IAEA currently uses the Cicero multichannel analyzer and the IAEAPU algorithm for its analysis. In our investigation the plutonium isotopic measurements were found to be good for PuO/sub 2/ powder or low-enriched MOX, but acceptable for highly enriched MOX in IAEA special nuclear material (SNM) accountability applications. The gamma-ray interferences from /sup 235/U resulted in underestimation of the isotopic composition of /sup 239/Pu and overestimation of all other plutonium isotopes. Samples with high /sup 240/Pu content were found to have significantly higher error in plutonium isotopic analyses of highly enriched MOX. Code modifications or use of calibration curves are necessary for plutonium isotopic analyses of highly enriched MOX in IAEA SNM accountability applications.

Research Organization:
Los Alamos National Lab. (LANL), Los Alamos, NM (United States)
DOE Contract Number:
W-7405-ENG-36
OSTI ID:
5162039
Report Number(s):
LA-10810-MS; ON: DE87000800
Country of Publication:
United States
Language:
English