Recent improvements to RC-line encoded position-sensitive proportional counters
Conference
·
OSTI ID:5161454
Continuing research on the principles of position encoding with RC lines has advanced the design of position-sensitive proportional counters (PSPCs) to meet the requirements for high count rates (>10/sup 5/ counts/sec) and good spatial resolution (>10/sup 4/ spatial elements) in small-angle scattering experiments with x rays and neutrons. Low-noise preamplifiers were developed with pole-zero cancellation in the feedback circuit and modular linear amplifiers with passive RCL shaping which, compared to previous designs, reduce output saturation at high count rates approx.20 times and shorten the position signal processing time to <1 ..mu..sec. Using resistive self-heating and propylene decomposition, the conductance and film thickness of pyrolytic-carbon anodes were increased to match the RC time-constants of line PSPCs with the linear amplifiers, to reduce the effects of electron erosion of the anodes at high count rates, and to improve the PSPC linearity. A large-area PSPC (480 x 780 mm/sup 2/) for low-energy (<60 keV) photon imaging was developed. The electronic uncertainty in the position measurement of 20-keV photons is <0.5 mm (fwhm) in both dimensions. The technology developed for this PSPC is applicable to the design of large-area detectors (>800 x 800 mm/sup 2/) for the measurement of small-angle scattering with neutrons. The method of electronic thickness discrimination was applied to change the effective thickness of an area PSPC from 12 to 2 cm whenever the molybdenum target of an x-ray generator was changed to a copper target. This thickness adjustment increased the signal-to-background ratio by a factor of approx.6 for the 8-keV photons from the copper target, while maintaining a >90% detection efficiency.
- Research Organization:
- Oak Ridge National Lab., Tenn. (USA)
- DOE Contract Number:
- W-7405-ENG-26
- OSTI ID:
- 5161454
- Report Number(s):
- CONF-771077-5
- Country of Publication:
- United States
- Language:
- English
Similar Records
RC-encoding for position-sensitive proportional counters using low-resistance, metal-wire anode and active load capacitance
Line-scanning proportional counter camera
Large-area proportional counter camera for the US National Small-Angle Neutron Scattering Facility
Journal Article
·
Wed Feb 28 23:00:00 EST 1979
· Rev. Sci. Instrum.; (United States)
·
OSTI ID:6456579
Line-scanning proportional counter camera
Journal Article
·
Tue Apr 01 00:00:00 EDT 1975
· IEEE Trans. Nucl. Sci., v. NS-22, no. 2, pp. 896-900
·
OSTI ID:4182725
Large-area proportional counter camera for the US National Small-Angle Neutron Scattering Facility
Conference
·
Mon Dec 31 23:00:00 EST 1979
·
OSTI ID:5804859
Related Subjects
440101* -- Radiation Instrumentation-- General Detectors or Monitors & Radiometric Instruments
46 INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY
AMPLIFIERS
ANODES
COUNTING RATES
ELECTRODES
ELECTRONIC EQUIPMENT
IONIZATION CHAMBERS
MEASURING INSTRUMENTS
MULTIWIRE PROPORTIONAL CHAMBERS
POSITION SENSITIVE DETECTORS
PREAMPLIFIERS
PROPORTIONAL COUNTERS
RADIATION DETECTORS
READOUT SYSTEMS
RESOLUTION
SCATTERING
SMALL ANGLE SCATTERING
SPATIAL RESOLUTION
46 INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY
AMPLIFIERS
ANODES
COUNTING RATES
ELECTRODES
ELECTRONIC EQUIPMENT
IONIZATION CHAMBERS
MEASURING INSTRUMENTS
MULTIWIRE PROPORTIONAL CHAMBERS
POSITION SENSITIVE DETECTORS
PREAMPLIFIERS
PROPORTIONAL COUNTERS
RADIATION DETECTORS
READOUT SYSTEMS
RESOLUTION
SCATTERING
SMALL ANGLE SCATTERING
SPATIAL RESOLUTION