Skip to main content
U.S. Department of Energy
Office of Scientific and Technical Information

Calibration of a monochromator/spectrometer system for the measurement of photoelectron angular distributions and branching ratios

Conference ·
OSTI ID:5143625
 [1];  [2];  [2];  [3];  [4]
  1. Los Alamos National Lab. (LANL), Los Alamos, NM (United States)
  2. National Bureau of Standards, Washington, DC (United States). Synchrotron Ultraviolet Radiation Facility
  3. Argonne National Lab. (ANL), Argonne, IL (United States)
  4. Daresbury Lab. (United Kingdom)

We describe the techniques used in calibrating a monochromator/spectrometer system for gas-phase photoelectron angular distribution and branching ratio measurements. We report a self-consistent set of values for the Ne 2p, Ar 3p, Kr 4p3/2 and 4p1/2, and Xe 5p3/2 and 5p1/2 photoelectron asymmetry parameters and for the Kr 4p3/2:4p1/2 and Xe 5p3/2:5p1/2 branching ratios for the energy regions from threshold to approximately 15 eV. 22 refs., 7 figs.

Research Organization:
Los Alamos National Lab., NM (USA)
Sponsoring Organization:
USDOE
DOE Contract Number:
W-7405-ENG-36
OSTI ID:
5143625
Report Number(s):
LA-UR--85-3113; CONF-850734--13; ON: DE86000789
Country of Publication:
United States
Language:
English