Reduction of polyatomic ion interferences in inductively coupled plasma mass spectrometry by cryogenic desolvation

Journal Article · · Analytical Chemistry (Washington)
Multiple desolvation steps at -80 {degrees}C for a continuous-flow ultrasonic nebulizer reduced the load of water and HCl on an argon inductively coupled plasma (ICP). Consequently, the oxide ratios MO{sup +} /M{sup +} for Ca, Mo, and La were 0.02-0.05%. Polyatomic ion interferences from ArO{sup +}, ClO{sup +}, and ArCl{sup +} were also reduced by several orders of magnitude. Measured isotope ratios for Ni, Cd, and Fe were in good agreement with accepted natural values, and the detection limits were 2-20 ng L{sup -1} for Ni, Fe, V, As, and Cd in the presence of concentrated matrices of HCl, Ca, or Mo that would normally cause polyatomic ion interference. Eight additional seconds were necessary for sample rinse-out, and analyte sensitivity was reduced by a factor of 2 when cryogenic desolvation was used. UO{sup +}/U{sup +} ratios were reduced to 0.06% when cryogenic desolvation was used and acetylene was added to the central channel. Sodium matrices up to 100 mg L{sup -1} decreased the La{sup +} count rate but did not affect the oxide ratio LaO{sup +}/La{sup +}. At 1000 mg L{sup -1} NaCl the LaO{sup +}/La{sup +} ratio increased to 0.08%. 35 refs., 6 figs., 6 tabs.
Sponsoring Organization:
USDOE
DOE Contract Number:
W-7405-ENG-82
OSTI ID:
513383
Journal Information:
Analytical Chemistry (Washington), Journal Name: Analytical Chemistry (Washington) Journal Issue: 10 Vol. 64; ISSN ANCHAM; ISSN 0003-2700
Country of Publication:
United States
Language:
English

Similar Records

Reduction of polyatomic ion interferences in inductively coupled plasma mass spectrometry with cryogenic desolvation
Thesis/Dissertation · Thu Dec 31 23:00:00 EST 1992 · OSTI ID:6981284

Measurement of vanadium, nickel, and arsenic in seawater and urine reference materials by inductively coupled plasma mass spectrometry with cryogenic desolvation
Journal Article · Wed Sep 15 00:00:00 EDT 1993 · Analytical Chemistry (Washington); (United States) · OSTI ID:6915324

Estimation of the formation rates of polyatomic species of heavy metals in plutonium analyses using a multicollector ICP-MS with a desolvating nebulizer
Journal Article · Wed Dec 31 23:00:00 EST 2014 · Journal of Analytical Atomic Spectrometry · OSTI ID:1188890