skip to main content
OSTI.GOV title logo U.S. Department of Energy
Office of Scientific and Technical Information

Title: Radio-frequency glow discharge ion trap mass spectrometry

Journal Article · · Analytical Chemistry (Washington)
;  [1]; ;  [2]
  1. Oak Ridge National Lab., TN (United States)
  2. Clemson Univ., SC (United States)

A radio-frequency glow discharge ion source has been interfaced with a three-dimensional quadrupole ion trap. Data are described for ion derived from brass and multicomponent glass standards. These data indicate that the ion trap has some interesting features as a mass analyzer for ions formed in a glow discharge. MS/MS efficiencies approaching 100% for oxides and hydroxides of all metal ions studied were observed. Ions derived from the argon support gas, such as ArH{sup +}, Ar{sup 2+}, and Ar{sub 2}{sup {sm_bullet}+}, were depleted in the ion trap by rapid electron-and proton-transfer reactions to background gases in the ion trap. The mass/charge ratios of the terminal ions in the sequence of reactions fell below the lower mass limit of the ion trap and were quickly ejected. The use of mass-selective ion injection techniques permitted the analysis of components present at levels as low as a few tens of parts per million. However, chemical noise due to relatively high background pressures in the ion source precluded analysis of components of lesser abundance. 35 refs., 6 figs.

Sponsoring Organization:
USDOE
DOE Contract Number:
AC05-84OR21400
OSTI ID:
513332
Journal Information:
Analytical Chemistry (Washington), Vol. 64, Issue 14; Other Information: PBD: 15 Jul 1992
Country of Publication:
United States
Language:
English