Skip to main content
U.S. Department of Energy
Office of Scientific and Technical Information

Estimated spectrum of a 4-MV therapeutic beam

Journal Article · · Med. Phys.; (United States)
DOI:https://doi.org/10.1118/1.595878· OSTI ID:5133113
We have made a comparative investigation of the estimated spectra obtained by the Laplace transform analysis of the transmitted exposure data measured in an absorption study of a 4-MV x-ray beam. Four transform pair models currently used with this method have been evaluated. It has been determined that the Archer-Wagner model provides a valid representation of the measured transmission data and yields an estimated spectrum which most closely resembles a Monte Carlo spectrum calculated for a 4-MV therapeutic x-ray beam available from a typical medical accelerator.
OSTI ID:
5133113
Journal Information:
Med. Phys.; (United States), Journal Name: Med. Phys.; (United States) Vol. 3; ISSN MPHYA
Country of Publication:
United States
Language:
English