Convergence and resolution recovery of block-iterative EM algorithms modeling 3D detector response in SPECT

Conference ·
OSTI ID:513263
; ;  [1]
  1. Univ. of North Carolina, Chapel Hill, NC (United States)

We evaluate fast reconstruction algorithms including ordered subsets-EM (OS-EM) and Rescaled Block Iterative EM (RBI-EM) in fully 3D SPECT applications on the basis of their convergence and resolution recovery properties as iterations proceed. Using a 3D computer-simulated phantom consisting of 3D Gaussian objects, we simulated projection data that includes only the effects of sampling and detector response of a parallel-hole collimator. Reconstructions were performed using each of the three algorithms (ML-EM, OS-EM, and RBI-EM) modeling the 3D detector response in the projection function. Resolution recovery was evaluated by fitting Gaussians to each of the four objects in the iterated image estimates at selected intervals. Results show that OS-EM and RBI-EM behave identically in this case; their resolution recovery results are virtually indistinguishable. Their resolution behavior appears to be very similar to that of ML-EM, but accelerated by a factor of twenty. For all three algorithms, smaller objects take more iterations to converge. Next, we consider the effect noise has on convergence. For both noise-free and noisy data, we evaluate the log likelihood function at each subiteration of OS-EM and RBI-EM, and at each iteration of ML-EM. With noisy data, both OS-EM and RBI-EM give results for which the log-likelihood function oscillates. Especially for 180-degree acquisitions, RBI-EM oscillates less than OS-EM. Both OS-EM and RBI-EM appear to converge to solutions, but not to the ML solution. We conclude that both OS-EM and RBI-EM can be effective algorithms for fully 3D SPECT reconstruction. Both recover resolution similarly to ML-EM, only more quickly.

OSTI ID:
513263
Report Number(s):
CONF-961123--; CNN: Grant CA39463
Country of Publication:
United States
Language:
English

Similar Records

Evaluation of OS-EM vs. ML-EM for 1D, 2D and fully 3D SPECT reconstruction
Journal Article · Sat Jun 01 00:00:00 EDT 1996 · IEEE Transactions on Nuclear Science · OSTI ID:267974

Region of interest evaluation of SPECT image reconstruction methods using a realistic brain phantom
Conference · Mon Dec 30 23:00:00 EST 1996 · OSTI ID:513249

Comparison between ML-EM and WLS-CG algorithms for SPECT image reconstruction
Journal Article · Sat Nov 30 23:00:00 EST 1991 · IEEE Transactions on Nuclear Science (Institute of Electrical and Electronics Engineers); (United States) · OSTI ID:5745622