Title: MicroPET: A high resolution PET scanner for imaging small animals

Conference ·
OSTI ID:513161
; ;  [1]
  1. Crump Institute for Biological Imaging, Los Angeles, CA (United States); and others

MicroPET is a high resolution positron emission tomography (PET) scanner designed for imaging small laboratory animals. It consists of a ring of 30 position-sensitive scintillation detectors, each with an 8 x 8 array of small lutetium oxyorthosilicate (LSO) crystals coupled via optical fibers to a multi-channel photomultiplier tube. The detectors have an intrinsic resolution averaging 1.68 mm, an energy resolution between 15 and 25% and 2.4 ns timing resolution at 511 keV. The detector ring diameter of microPET is 17.2 cm with an imaging field of view of 112 mm transaxially by 18 mm axially. The scanner has no septa and operates exclusively in 3D mode. Reconstructed image resolution 1 cm from the center of the scanner is 2.0 mm and virtually isotropic, yielding a volume resolution of 8 mm{sup 3}. For comparison, the volume resolution of state-of-the-art clinical PET systems is in the range of 50 - 75 mm{sup 3}. Initial images of phantoms have been acquired and are reported. A computer controlled bed is under construction and will incorporate a small wobble motion to improve spatial sampling. This is projected to further enhance spatial resolution. MicroPET is the first PET scanner to incorporate the new scintillator LSO and to our knowledge is the highest resolution multi-ring PET scanner currently in existence.

DOE Contract Number:
FC03-87ER60615
OSTI ID:
513161
Report Number(s):
CONF-961123-; CNN: Grant J100193; Grant R01 CA69370; TRN: 97:014214
Resource Relation:
Conference: Institute of Electrical and Electronic Engineers (IEEE) nuclear science symposium and medical imaging conference, Anaheim, CA (United States), 2-9 Nov 1996; Other Information: PBD: 1996; Related Information: Is Part Of 1996 IEEE nuclear science symposium - conference record. Volumes 1, 2 and 3; Del Guerra, A. [ed.]; PB: 2138 p.
Country of Publication:
United States
Language:
English

Similar Records

Studies with a prototype high resolution PET scanner based on LSO-APD modules
Journal Article · Mon Jun 01 00:00:00 EDT 1998 · IEEE Transactions on Nuclear Science · OSTI ID:513161

Evaluation of cerium doped lutetium oxyorthosilicate (LSO) scintillation crystal for PET
Journal Article · Sun Aug 01 00:00:00 EDT 1993 · IEEE Transactions on Nuclear Science (Institute of Electrical and Electronics Engineers); (United States) · OSTI ID:513161

Effects of reflector and crystal surface on the performance of a depth-encoding PET detector with dual-ended readout
Journal Article · Tue Jul 15 00:00:00 EDT 2014 · Medical Physics · OSTI ID:513161