Improvement of the skin sparing characteristics of the clinac 4 by the use of leaded glass electron filters

Journal Article · · Int. J. Radiat. Oncol., Biol. Phys.; (United States)
Ionization chamber measurements were taken to determine the Relative Skin Dose from the Varian Clinac 4 (lead flattening filter). The effect of blocking tray material, field size, and scatterer-skin distance on the relative skin dose was investigated both on and off the central axis. It was found that the leaded glass tray enhanced the skin sparing effect of the beam by 25 to 30 percent.
Research Organization:
Univ. of North Carolina, Chapel Hill
OSTI ID:
5089831
Journal Information:
Int. J. Radiat. Oncol., Biol. Phys.; (United States), Journal Name: Int. J. Radiat. Oncol., Biol. Phys.; (United States) Vol. 2:9/10; ISSN IOBPD
Country of Publication:
United States
Language:
English