Skip to main content
U.S. Department of Energy
Office of Scientific and Technical Information

Neutron detection based on superheated materials

Conference ·
DOI:https://doi.org/10.1109/23.106734· OSTI ID:5078084
; ;  [1];  [2]
  1. Pacific Northwest Lab., Richland, WA (USA)
  2. Apfel Enterprises, Inc. (USA)
The environmental and radiation responses of the Active Personnel Dosimeter/Superheated Drop Detector (APD/SDD) combination have been evaluated at the Pacific Northwest Laboratory (PNL) for the US Department of Energy's Neutron Measurement and Evaluation Program. This paper provides results of the evaluation and discusses possible improvements for the current system. Radiation detection based on the radiation sensitivity of superheated liquids has been studied for some time. A liquid is superheated if it exists as a liquid at a temperature-pressure state normally associated with the vapor phase of that material. The liquid does not vaporize because there are no bubble nucleation sites in the sample. These sites usually exist (1) in microscopic cracks on solid container surfaces, (2) in crevices of imperfectly wetted solid particles suspended in the liquid, or (3) as a result of the radiation-matter interaction producing a microbubble that is large enough for bubble growth to be thermodynamically favorable. By suspending small drops of superheated liquid in an immiscible, inert, impurity-free medium, potential for bubble nucleation by the first two mechanisms is eliminated. Therefore, each drop becomes a miniature radiation detector.
Research Organization:
Pacific Northwest Lab., Richland, WA (USA)
Sponsoring Organization:
DOE/EH
DOE Contract Number:
AC06-76RL01830
OSTI ID:
5078084
Report Number(s):
PNL-SA-17137; CONF-900143--14; ON: DE90006124
Country of Publication:
United States
Language:
English

Similar Records

Neutron detection based on superheated materials
Conference · Sat Mar 31 23:00:00 EST 1990 · IEEE Transactions on Nuclear Science (Institute of Electrical and Electronics Engineers); (USA) · OSTI ID:6872098

The Active Personnel Dosimeter: Apfel Enterprises superheated drop detector
Conference · Sun Oct 01 00:00:00 EDT 1989 · OSTI ID:5303045

The active personnel dosimeter---APFEL enterprises superheated drop detector
Conference · Thu Feb 28 23:00:00 EST 1991 · OSTI ID:6042205