Skip to main content
U.S. Department of Energy
Office of Scientific and Technical Information

Preliminary study of a gamma-ray telescope relevant to neutron personnel dosimetry

Technical Report ·
OSTI ID:5018045
The objective was to use a gamma-ray dosemeter worn on the body to measure the neutron dose equivalent in the energy range <10 keV. It has been shown for cobalt 60 gamma rays that the reading of a 0.4 mm-thick LiF thermoluminescent dosemeter is only about half as large if irradiated just behind a Pb layer as if irradiated just in front of it. This report examines the possibility of using the ratio of TLD readings from dosemeters on opposite sides of a Pb filter as a measure of the relative dose contributions of gamma rays travelling in opposite directions. (ACR)
Research Organization:
Wisconsin Univ., Madison (USA). Dept. of Medical Physics
DOE Contract Number:
AC02-76EV01105
OSTI ID:
5018045
Report Number(s):
DOE/EV/01105-289; ON: DE82019976
Country of Publication:
United States
Language:
English