Skip to main content
U.S. Department of Energy
Office of Scientific and Technical Information

Lead shielding thickness for dose reduction of 5 MeV electrons

Journal Article · · Medical Physics; (USA)
DOI:https://doi.org/10.1118/1.596432· OSTI ID:5013356
; ; ;  [1]
  1. Northwestern Univ. Medical School, IL (USA)
The relative percent intensity reduction by lead (Pb) of 5 MeV electrons produced by Siemens Mevatron 77/74 for 5 cm diameter, 10 X 10, 15 X 15, and 20 X 20 cm2 cones both with and without buildup is measured. The thickness of lead (Pb) required to attenuate the intensity of the primary electron beam to 95% and 98% depends upon the cone size and upon the depth in phantom at which transmission measurements are made.
OSTI ID:
5013356
Journal Information:
Medical Physics; (USA), Journal Name: Medical Physics; (USA) Vol. 16:5; ISSN 0094-2405; ISSN MPHYA
Country of Publication:
United States
Language:
English

Similar Records

Lead shielding thickness for dose reduction of 6-MeV electrons for different square fields
Journal Article · Mon Feb 29 23:00:00 EST 1988 · Med. Phys.; (United States) · OSTI ID:5421066

Lead shielding thickness for dose reduction of 7- to 28-MeV electrons
Journal Article · Fri Oct 31 23:00:00 EST 1975 · Med. Phys., v. 2, no. 6, pp. 336-337 · OSTI ID:4131369

Lipowitz metal shielding thickness for dose reduction of 6--20 MeV electrons