Two-color infrared thermometer for low-temperature measurement using a hollow glass optical fiber

Conference ·
OSTI ID:495673

In the thermometer, radiation from a target is collected via a single 700 {mu}m-bore hollow glass optical fiber coated with a metallic/dielectric layer on the inner surface, simultaneously split into two paths and modulated by a Au-coated reflective chopper, and focused onto two thermoelectrically cooled mid-infrared HgCdZnTe photoconductors by 128.8 mm-radius Au-coated spherical mirrors. The photoconductors have spectral bandpasses of 2-6 {mu}m and 2.12 {mu}m, respectively. The modulated detector signals are recovered using lock- in amplification. The two signals are calibrated using a blackbody (emissivity=1) of known temperature, and exponential fits are applied to the two resulting voltage vs temperature curves. Using the two calibration equations, a computer algorithm calculates the temperature and emissivity of a target in real time, taking into account reflection of the background radiation field from the target surface.

Research Organization:
Lawrence Livermore National Lab., CA (United States)
Sponsoring Organization:
USDOE, Washington, DC (United States)
DOE Contract Number:
W-7405-ENG-48
OSTI ID:
495673
Report Number(s):
UCRL-JC--126840; CONF-970231--42; ON: DE97053190; CRN: C/LLNL--TC-1085-94
Country of Publication:
United States
Language:
English

Similar Records

Two-color mid-infrared thermometer using a hollow glass optical fiber
Conference · Mon Jun 30 00:00:00 EDT 1997 · OSTI ID:304678

Two-color mid-infrared thermometer with a hollow glass optical fiber
Journal Article · Thu Oct 01 00:00:00 EDT 1998 · Applied Optics · OSTI ID:675102

Single-fiber multi-color pyrometry
Patent · Fri Dec 31 23:00:00 EST 1999 · OSTI ID:872804