GAMMA-RAY DISCRIMINATION IN A PROTON-RECOIL PROPORTIONAL COUNTER

Technical Report ·
OSTI ID:4794110
A technique for discriminating against gamma background in proton-recoil proportional counters based on changes in the pulse profile is described. The lower useful energy limit of a counter filled with a propane-nitrogen gas mixture and placed in a broad fast-neutron spectrum was extended from 400 to 100 kev. (auth)
Research Organization:
Argonne National Lab., Ill.
DOE Contract Number:
W-31109-ENG-38
NSA Number:
NSA-16-014900
OSTI ID:
4794110
Report Number(s):
ANL-6480
Country of Publication:
United States
Language:
English

Similar Records

Proportional Counter Proton-Recoil Spectrometer with Gamma Discrimination
Journal Article · Wed Oct 31 23:00:00 EST 1962 · Review of Scientific Instruments · OSTI ID:4735605

Proton-recoil proportional counter tests at TREAT
Conference · Sun Dec 31 23:00:00 EST 1978 · OSTI ID:5713784

Proton-recoil proportional-counter array for neutron-image construction
Conference · Sat Dec 31 23:00:00 EST 1983 · OSTI ID:6611343