Skip to main content
U.S. Department of Energy
Office of Scientific and Technical Information

A LOW-NOISE CHARGE-SENSITIVE PREAMPLIFIER FOR SEMI-CONDUCTOR IONIZATION CHAMBERS (in German)

Journal Article · · Atomkernenergie (West Germany) Merged with Kerntechnik to form Atomkernenerg./Kerntech. Acta Radiol. Changed to Acta Radiol.: Oncol., Radiat. Phys.
OSTI ID:4760834
A low-noise charge-sensitive preamplifier for particle spectroscopy with junction detectors is described. Using an E 280 F as input tube a noise width of 2.9 kev (fwhm) is achieved. The open-loop gain is in excess of 1800. The noise width as a function of the external input capacity is measured. Without a feedback capacitor the circuit may be used as a voltage-sensitive preamplifier with an extremely good noise performance. (auth)
Research Organization:
Gesellschaft fur Kernenergieverwertung in Schiffbau und Schiffahrt mbh, Ger.
NSA Number:
NSA-17-003318
OSTI ID:
4760834
Journal Information:
Atomkernenergie (West Germany) Merged with Kerntechnik to form Atomkernenerg./Kerntech. Acta Radiol. Changed to Acta Radiol.: Oncol., Radiat. Phys., Journal Name: Atomkernenergie (West Germany) Merged with Kerntechnik to form Atomkernenerg./Kerntech. Acta Radiol. Changed to Acta Radiol.: Oncol., Radiat. Phys. Vol. Vol: 7; ISSN ATKEA
Country of Publication:
Country unknown/Code not available