PLUTONIUM ALPHA AIR MONITOR USING A SOLID STATE DETECTOR
Technical Report
·
OSTI ID:4714129
A continuously indicating Pu alpha air monitor using a P--N junction Si detector and single-channel pulse-height analyzer is described. The resolution of the system excludes 98% of the radon daughter aipha activity. One MPC of Pu/ sup 238/ can easily be seen within 8 hours of sampling. (auth)
- Research Organization:
- California. Univ., Livermore. Lawrence Radiation Lab.
- DOE Contract Number:
- W-7405-ENG-48
- NSA Number:
- NSA-17-011012
- OSTI ID:
- 4714129
- Report Number(s):
- UCRL-6692
- Country of Publication:
- United States
- Language:
- English
Similar Records