Absorbed Dose Calibration of an Ion Chamber
A method for calibrating an ion chamber in terms of absorbed dose is given which avoids the need for corrections for attenuation. The absorbed dose is measured using a carbon calorimeter, a carbon chamber, and a BaldwinFarmer chamber. The method was used with x-ray beams at 4 to 20 Mv and Co⁶⁰ gamma rays. The mean energy per ion pair in air at 0 deg C 760 mm Hg was determined to be 34.11 plus or minus 0.21 ev in carbon. A graph is given for the rad calibration factor of the Baldwin- Farmer chamber exposed in Perspex. (D.L.C.)
- Research Organization:
- Hammersmith Hospital, London
- Sponsoring Organization:
- USDOE
- NSA Number:
- NSA-17-029214
- OSTI ID:
- 4702163
- Journal Information:
- Nature (London), Journal Name: Nature (London) Journal Issue: 4887 Vol. 198; ISSN 0028-0836
- Publisher:
- Nature Publishing Group
- Country of Publication:
- Country unknown/Code not available
- Language:
- English
Similar Records
Exposure calibration of a thimble chamber by calorimetric and ionimetric methods
ON THE USE OF ROENTGEN-CALIBRATED, CAVITY-IONIZATION CHAMBERS IN TISSUE- LIKE PHANTOMS TO DETERMINE ABSORBED DOSE
Graphite calorimeter in water phantom and calibration of ionization chambers in dose to water for /sup 60/Co gamma radiation
Technical Report
·
Wed Aug 01 00:00:00 EDT 1973
·
OSTI ID:4935291
ON THE USE OF ROENTGEN-CALIBRATED, CAVITY-IONIZATION CHAMBERS IN TISSUE- LIKE PHANTOMS TO DETERMINE ABSORBED DOSE
Journal Article
·
Mon Oct 01 00:00:00 EDT 1962
· Physics in Medicine and Biology (England)
·
OSTI ID:4773177
Graphite calorimeter in water phantom and calibration of ionization chambers in dose to water for /sup 60/Co gamma radiation
Journal Article
·
Thu May 01 00:00:00 EDT 1980
· Med. Phys.; (United States)
·
OSTI ID:5429847