Skip to main content
U.S. Department of Energy
Office of Scientific and Technical Information

INERT GAS FUSION DETERMINATION OF OXYGEN IN VANADIUM, NIOBIUM, AND TANTALUM

Journal Article · · Analytical Chemistry (U.S.) Formerly Ind. Eng. Chem., Anal. Ed.
OSTI ID:4696940

>The inert gas fusion technique was applied to the determination of oxygen in the Group V-A metals---vanadium, niobium, and tantalum. A platinum flux technique was satisfactory for the determination of oxygen in vanadium and a mixed nickel--platinum bath used with a platinum flux technique was satisfactory for niobium and tantalum. Oxygen contents from 100 to 1000 ppm were determined on standard samples with relative standard deviation of 3% for vanadium and 5% for niobium and tantalum. (auth)

Research Organization:
Iowa State Univ., Ames
NSA Number:
NSA-17-027257
OSTI ID:
4696940
Report Number(s):
IS-614; 0003-2700
Journal Information:
Analytical Chemistry (U.S.) Formerly Ind. Eng. Chem., Anal. Ed., Journal Name: Analytical Chemistry (U.S.) Formerly Ind. Eng. Chem., Anal. Ed. Vol. Vol: 35; ISSN ANCHA
Country of Publication:
Country unknown/Code not available
Language:
English

Similar Records

A COMPARISON OF THE INERT GAS-FUSION AND VACUUM EXTRACTION METHODS FOR THE DETERMINATION OF OXYGEN IN REFRACTORY METALS
Technical Report · Thu Dec 31 23:00:00 EST 1959 · OSTI ID:4791513

Determination of Oxygen in Mixed Fluorides by Inert Gas Fusion.
Journal Article · Wed Oct 31 23:00:00 EST 1962 · Analytical Chemistry · OSTI ID:4749802

VACUUM-FUSION ANALYSIS FOR OXYGEN IN TITANIUM
Technical Report · Sat Nov 30 23:00:00 EST 1957 · OSTI ID:4318332