Automatic, Amperometric, Cupferron Titration of Zirconium in Highly Radioactive Solutions

Journal Article · · Analytical Chemistry
DOI:https://doi.org/10.1021/ac60204a052· OSTI ID:4628841
A previously reported amperometric titration of zirconium and cupferron was modified and adapted for hot cell analysis. A titration was devised in which the current is monitored continuously while the titrant is fed in at a constant rate. A stirred mercury pool is the indicating electrode that senses the point where cupferron comes into excess, and a platinum wire is used for the reference. Precision of 1% relative standard deviation is obtainable for benchtop analysis and 1.5% for titrations in a hot cell. The titration was specifically devised for the analysis of zirconium in molten salt reactor fuel, and the titration can be performed directly following dissolution without the need for separations.
Research Organization:
Oak Ridge National Lab., Tenn.
Sponsoring Organization:
USDOE
NSA Number:
NSA-17-038930
OSTI ID:
4628841
Journal Information:
Analytical Chemistry, Journal Name: Analytical Chemistry Journal Issue: 11 Vol. 35; ISSN 0003-2700
Publisher:
American Chemical Society (ACS)
Country of Publication:
Country unknown/Code not available
Language:
English

Similar Records

DETERMINATION OF ZIRCONIUM BY AMPEROMETRIC TITRATION BY APPLICATION OF CUPFERRON
Journal Article · Mon Dec 31 23:00:00 EST 1956 · Zavodskaya Lab. · OSTI ID:4290468

AMPEROMETRIC DETERMINATION OF TIN WITH CUPFERRON
Technical Report · Mon Mar 31 23:00:00 EST 1958 · OSTI ID:4315990

AMPEROMETRIC TITRATION OF THORIUM SALTS WITH CUPFERRON
Journal Article · Sat Jul 01 00:00:00 EDT 1961 · Zhur. Anal. Khim. · OSTI ID:4838179