Title: Signal processing and general purpose data acquisition system for on-line tomographic measurements

Journal Article · · Review of Scientific Instruments
DOI:https://doi.org/10.1063/1.1147723· OSTI ID:451925
; ; ; ; ;  [1];  [2]
  1. Padova Group for Fusion Research, Euratom-Enea-Cnr-Universita di Padova Associations, Corso Stati Uniti 4, 35127 Padova (Italy)
  2. Physikalisch-Technische Studien GmbH, Leinenweberstrasse 16, D-79108 Freiburg (Germany)

New analog signal conditioning electronics and data acquisition systems have been developed for the soft x-ray and bolometric tomography diagnostic in the reverse field pinch experiment (RFX). For the soft x-ray detectors the analog signal processing includes a fully differential current to voltage conversion, with up to a 200 kHz bandwidth. For the bolometers, a 50 kHz carrier frequency amplifier allows a maximum bandwidth of 10 kHz. In both cases the analog signals are digitized with a 1 MHz sampling rate close to the diagnostic and are transmitted via a transparent asynchronous xmitter/receiver interface (TAXI) link to purpose built Versa Module Europa (VME) modules which perform data acquisition. A software library has been developed for data preprocessing and tomographic reconstruction. It has been written in C language and is self-contained, i.e., no additional mathematical library is required. The package is therefore platform-free: in particular it can perform online analysis in a real-time application, such as continuous display and feedback, and is portable for long duration fusion or other physical experiments. Due to the modular organization of the library, new preprocessing and analysis modules can be easily integrated in the environment. This software is implemented in RFX over three different platforms: open VMS, digital Unix, and VME 68040 CPU. {copyright} {ital 1997 American Institute of Physics.}

OSTI ID:
451925
Report Number(s):
CONF-960543-; ISSN 0034-6748; TRN: 97:006282
Journal Information:
Review of Scientific Instruments, Vol. 68, Issue 1; Conference: 11. annual high-temperature plasma diagnostics conference, Monterey, CA (United States), 12-16 May 1996; Other Information: PBD: Jan 1997
Country of Publication:
United States
Language:
English

Similar Records

Soft x-ray and bolometric tomography in RFX
Journal Article · Sat Feb 01 00:00:00 EST 1997 · Review of Scientific Instruments · OSTI ID:451925

Soft x-ray and bolometric tomography in RFX (abstract){sup a}
Journal Article · Wed Jan 01 00:00:00 EST 1997 · Review of Scientific Instruments · OSTI ID:451925

DATA ACQUISITION FOR SNS BEAM LOSS MONITOR SYSTEM
Conference · Mon Oct 13 00:00:00 EDT 2003 · OSTI ID:451925