Skip to main content
U.S. Department of Energy
Office of Scientific and Technical Information

Electron and x-ray beam characteristics of the linac LMR-13

Journal Article · · Radiology, v. 110, no. 1, pp. 203-207
OSTI ID:4384797
Electron beams of nominal energies 4, 8, 10, and l2 MeV were investigated. The effect of Ni and Pb scattering folla on beam penetration and field flatness is discussed and optimal combinations of beam energies and scattering foils are suggested. Depth doses are expressed as percentiles. A study of the 10 MV x-ray beam shows a need for two flattening filters for clinical ranges of field sizes. Flatness and depth dose data variations for two lead flattening filters of the same maximum thickness and central axis percentage depth doses are given. (auth)
Research Organization:
Univ. of Miami, FL
NSA Number:
NSA-29-016454
OSTI ID:
4384797
Journal Information:
Radiology, v. 110, no. 1, pp. 203-207, Journal Name: Radiology, v. 110, no. 1, pp. 203-207; ISSN RADLA
Country of Publication:
Country unknown/Code not available
Language:
English