Measurement of the dead-time in a multichannel analyser
Journal Article
·
· Nuclear Instruments and Methods
By means of two simple measurements three different deadtimes are determined: the normal dead-time, a dead-time coming from the pile-up, and a dead- time due to the finite width of the timing pulses.