The tomography beamline at the National Synchrotron Light Source
- and others
We compared the image contrast of a monochromatic CT, Multiple Energy Computed Tomography (MECT), and conventional CT scanner using phantoms. The experimental results indicate that monochromatic CT, with beam energy tuned just above the iodine K-edge, has about a 3 fold advantage in iodine contrast over conventional CT with a 120 kVp beam. Modeling using the same beams at a 3 rad dose and 3 mm slice height on an 18 cm diameter acrylic phantom, the simulations show a noise of 1.2 HU for MECT and 1.9 HU for CCT. Furthermore, despite the Cupping-effect corrections the bone contrast is lower in CCT and varies by 24 HU moving from the phantom`s center to the edge; this indicates an advantage for MECT in detecting and quantifying lesions differing from surrounding tissue by their mean atomic number.