Characteristics of a neutron moisture gage with a solid state detector

Journal Article · · Pramana, v. 1, no. 2, pp. 112-116
OSTI ID:4346020
The development of a neutron moisture gauge, using /sup 252/Cf as a source of fast neutrons and muscovite as a detector of thermalneutron-induced- fission in /sup 239/Pu target, is reported. The laboratory and the field calibrations of the instrument reveal a linear relation between track count rate and the moisture volume fraction. Low cost, low weight, simplicity of operation, thermal stability and elimination of electronic components and power supplies are the attractive features of this instrument. The main drawback is its poor detection efficiency for thermal neutrons and the consequent unsuitability as a routine logging instrument. Its special features make it particularly suited to deep borehole logging such as in oil exploration. (auth)
Research Organization:
Kurukshetra Univ., India
NSA Number:
NSA-29-018509
OSTI ID:
4346020
Journal Information:
Pramana, v. 1, no. 2, pp. 112-116, Journal Name: Pramana, v. 1, no. 2, pp. 112-116; ISSN PRAMC
Country of Publication:
Country unknown/Code not available
Language:
English