Twin-tube conduction radionuclide calorimeter

Technical Report ·
DOI:https://doi.org/10.2172/4329039· OSTI ID:4329039
The complete calorimeter system used to measure powers from 0.1 W to 2 W of samples 1.5 in. in diameter by 5 in. in length is described. Calorimeter readout is by digital voltmeter. A sensitivity of 1501 mV/W at 1.327 W with on accuracy of 99.9% was obtained. Sample equilibrium time is 4 h. The twin-tube temperature measurement system compensates for thermostat temperature fiuctuations. Thermistors measure the temperature gradient produced by heat flow through a 5-in. length of 1.75-in.-o.d., 0.035-in.-wall brass tube. The thermistors are supplied with a constant operating current by current regulator diodes to simplify and improve operation. sn integral calibration heater surrounds the sample chamber. Reliability is enhanced by all-solid-state design. (auth)