Determination of Uranium Dioxide in Stainless Steels by the X-Ray Fluorescence Method
- North American Aviation, Inc., Canoga Park, CA (United States). Atomics International Div.
This paper describes a rapid method for the determination of uranium dioxide in stainless steel by direct X-ray fluorescent analysis after chemical solution of the sample in perchloric acid. Strontium is used as an internal standard. X-ray fluorescence methods of analysis have many advantages over the commonly used chemical methods such as the colorimetric, gravimetric and volumetric, because elements which ordinarily interfere need not be separated before analysis. Statistically, eleven independent solutions were counted and nine sets of results were obtained on each one. Standard deviations are shown.
- Research Organization:
- North American Aviation, Inc., Canoga Park, CA (United States). Atomics International Div.
- Sponsoring Organization:
- US Atomic Energy Commission (AEC)
- DOE Contract Number:
- AT(11-1)-GEN-8
- NSA Number:
- NSA-11-006257
- OSTI ID:
- 4328920
- Report Number(s):
- NAA-SR--1848
- Country of Publication:
- United States
- Language:
- English
Similar Records
X-RAY FLUORESCENCE ANALYSIS OF STAINLESS STEEL IN AQUEOUS SOLUTIONS
Preparation of uranium standard solutions for x-ray fluorescence analysis
THE VOLUMETRIC DETERMINATION OF URANIUM AFTER REDUCTION BY LEAD IN DILUTE PERCHLORIC ACID SOLUTION
Journal Article
·
Thu Oct 01 00:00:00 EDT 1959
· Analytical Chemistry (U.S.) Formerly Ind. Eng. Chem., Anal. Ed.
·
OSTI ID:4224959
Preparation of uranium standard solutions for x-ray fluorescence analysis
Conference
·
Tue Feb 28 23:00:00 EST 1978
·
OSTI ID:6805496
THE VOLUMETRIC DETERMINATION OF URANIUM AFTER REDUCTION BY LEAD IN DILUTE PERCHLORIC ACID SOLUTION
Journal Article
·
Thu Dec 31 23:00:00 EST 1959
· Analyst
·
OSTI ID:4187510