Skip to main content
U.S. Department of Energy
Office of Scientific and Technical Information

ON THE MEASUREMENT OF TRANSIT TIME DISPERSION IN MULTIPLIER PHOTOTUBES

Journal Article · · IRE Trans. Nuclear Sci.
In the use of a multiplier phototube to measure short time intervals, the accuracy of the measurement is limited by the statistical deviations in the output pulse from the phototube. These deviations arise from variations in the electron transit time in the phototube. This paper reports a technique by which the transit time dispersion in the tube structure can be measured. A beam of electrons is deflected through an aperture onto the first dynode of the photomultiplier by an rf sweep voltage. The output pulse resulting is examined using the sweep voltage as a time base. Values of the order of 1 m mu sec were obtained using three commercial type photomultiplier tubes. (auth)
Research Organization:
RCA Labs., Princeton, N.J.
Sponsoring Organization:
USDOE
NSA Number:
NSA-12-010764
OSTI ID:
4317869
Journal Information:
IRE Trans. Nuclear Sci., Journal Name: IRE Trans. Nuclear Sci. Vol. Vol: NS-5, No. 1
Country of Publication:
Country unknown/Code not available
Language:
English

Similar Records

SPECIAL PURPOSE MULTIPLIER PHOTOTUBES FOR SCINTILLATION COUNTING
Journal Article · Sun Nov 30 23:00:00 EST 1958 · IRE Trans. Nuclear Sci. · OSTI ID:4285503

MULTIPLIER-PHOTOTUBE DEVELOPMENT PROGRAM AT RCA-LANCASTER
Journal Article · Wed Jun 01 00:00:00 EDT 1960 · IRE Trans. Nuclear Sci. · OSTI ID:4140578

High-rate phototube base
Conference · Mon Jan 31 23:00:00 EST 1977 · IEEE Trans. Nucl. Sci.; (United States) · OSTI ID:7107365